MST
ACM
100 Technical Overview |
|
          |
|
|
Performance
Specifications |
| |
|
Analysis Method |
Continuous
scan FTIR (Fourier transform infrared) analyzer |
|
Gas cell path |
6 m or 10
cm path |
|
Gases monitored |
Organic,
PFC, CFC, HFC, Metal Organic, NF3 and many other
inorganic gases |
|
Lower detection limit |
~1.0 ppm
for 6 m path cell |
|
Scanning rate |
15 sec per
point |
|
Sample point capacity |
Available
in 20 and 40 point configurations |
|
Data archiving |
30 days
for all spectra
90 days for alarms |
|
Applicable standards |
S2-93
compliant, CE |
|
| |
| |
|
Sample line tubing |
1/4" ID x
3/8" OD polypropylene |
|
Sample line inlets |
3/8"
Swagelok® |
|
Electrical supply |
230 VAC,
10 A or 110 VAC, 20 A |
|
Purge gas, N2 |
5 - 10
psi, 2 l/min |
|
Venturi requirements |
90 - 120
psi, 320 l/min |
|
Exhaust pipe |
1, 5" NPT |
Overall dimensions
H x W x D |
66 x 34 x
25 in
(1,676 x 864 x 635mm) |
|
Weight operating |
700 lbs
(318 kg) |
|
Weight shipping |
930 lbs
(420 kg) |
|
| |
|
Outputs and
Communication |
| |
| Standard |
Relay outputs
|
80 Relays,
programmable by sample point and gas type |
Interfaces
|
Web based
user interface (e-diagnostics) |
|
| |
| Optional |
Interface
|
LonWorks®
Allen Bradley®, Modbus®, Profibus®
OPC drivers
others on request |
|
| |
|
| |
|
Analog output |
up to 40
(4 - 20 mA) |
|
Automatic sample line leak
test |
user
programmable |
|
|
|
|
|
|