MST Technology Website

Europe



Industrial Gas Detection; FTIR Hazardous Gas Monitor Europe

MST ACM 100 Technical Overview

 

 

Performance Specifications

 
Analysis Method Continuous scan FTIR (Fourier transform infrared) analyzer
Gas cell path 6 m or 10 cm path
Gases monitored Organic, PFC, CFC, HFC, Metal Organic, NF3 and many other inorganic gases
Lower detection limit ~1.0 ppm for 6 m path cell
Scanning rate 15 sec per point
Sample point capacity Available in 20 and 40 point configurations
Data archiving 30 days for all spectra
90 days for alarms
Applicable standards S2-93 compliant, CE
 

Facility Requirements

 
Sample line tubing 1/4" ID x 3/8" OD polypropylene
Sample line inlets 3/8" Swagelok®
Electrical supply 230 VAC, 10 A or 110 VAC, 20 A
Purge gas, N2 5 - 10 psi, 2 l/min
Venturi requirements 90 - 120 psi, 320 l/min
Exhaust pipe 1, 5" NPT
Overall dimensions
H x W x D
66 x 34 x 25 in
(1,676 x 864 x 635mm)
Weight operating 700 lbs (318 kg)
Weight shipping 930 lbs (420 kg)
 

Outputs and Communication

 
Standard

Relay outputs

80 Relays, programmable by sample point and gas type

Interfaces

Web based user interface (e-diagnostics)
 
Optional

Interface

LonWorks®
Allen Bradley®, Modbus®, Profibus
®
OPC drivers
others on request
 

Options and Accessories

 
Analog output up to 40 (4 - 20 mA)
Automatic sample line leak test user programmable
 
 
 
 

Industrial Gas Detection; FTIR Hazardous Gas Monitor Europe

Contact    Impressum    Add to favorites     Site Suggestions   Recommend Us   Privacy Statement
MST Technology - International Sites:  Global   
Europe     North America     Asia

Honeywell International    Honeywell Analytics


© MST Technology, Inc. 2005-2008 - All rights reserved

ISO 9001: 2000